##plugins.themes.bootstrap3.article.main##

The main purpose of QC program in diagnostic imaging modalities such as medical ultrasound equipment is to evaluate and define the performance and safety of ultrasound equipment. Ultrasound systems contain many sensitive transducers and soft parts which may be destroyed due to poor handling or use. Over time, any machines, regardless of their quality and usage, depreciates and incurs damage in many ways.

The results from the Quality Control (QC) reports for B-mode of the Ultrasound machines in the Clinical Imaging Department of Hamad Medical Corporation (HMC) has been reviewed. The average annual failures of equipment during the 4-year period includes 2.7% involved transducers and 51.3% involved mechanical integrity failures while the average annual failures rate from mechanical integrity and transducers were measured to be 12.8% and 0.7%, respectively. The main failure of the transducers could be from some artifacts or bad connection of individual elements or groups of elements. Evaluation of maximum depth of penetration showed that the measurement of distance accuracy is not effective in detecting equipment failures [1].

Downloads

Download data is not yet available.

References

  1. Hangiandreou NJ, Stekel SF, Tradup DJ, Gorny KR, King DM. Four-year experience with a clinical ultrasound quality control program. Ultrasound in Medicine & Biology. 2011
    DOI  |   Google Scholar
  2. American College of Radiology. Technical standard for diagnostic medical physics performance monitoring of real time ultrasound equipment. Reston, VA: American College of Radiology. 2011.
     Google Scholar
  3. American College of Radiology. Ultrasound accreditation program requirements. Reston, VA: American College of Radiology. 2015.
     Google Scholar
  4. American Institute of Ultrasound in Medicine. Routine quality assurance for diagnostic ultrasound equipment. Laurel, MD: American Institute of Ultrasound in Medicine. 2008.
     Google Scholar
  5. Goodsitt MM, Carson PL, Witt S, Hykes DL, Jr Kofler JM. Real-time B-mode ultrasound quality control test procedures. Report of AAPM Ultrasound Task Group No. 1. Med Phys. 1998; 25: 1385–1406.
    DOI  |   Google Scholar
  6. S Russell, (ed). Report no. 102. Quality assurance of ultrasound imaging systems. York, England: Institute of Physics and Engineering in Medicine; 2010.
     Google Scholar
  7. Institute of Physics and Engineering in Medicine (IPEM) Quality assurance of ultrasound imaging systems, report no. 102. York. 2009.
     Google Scholar
  8. Dudley N, Russell S, Ward B, Hoskins P. The BMUS guidelines for regular quality assurance testing of ultrasound scanners. Ultrasound. 2014.
    DOI  |   Google Scholar
  9. Sipila O. Quality assurance in diagnostic ultrasound. European Journal of Radiology. 2011; 80(2): 519-525.
    DOI  |   Google Scholar