Analysis of Sensitometry in Radiographic Films using Optical Density Measurement
Article Main Content
The sensitivity of radiographic films is an important factor to the clarity and accuracy of X-ray exposure to patients during treatment or diagnostic periods. It is therefore important to do a thorough analysis of the sensitivity of the radiographic film before and after exposure to enhance the Quality Assurance (QA) and the Quality Control (QC), of the exposure procedures. The optical densities (OD) of each film was measured, with a densitometer model MA 5336, made by GAMMEX. These values were then converted to the absorbed dose (X mGy), which is the amount of dose absorbed by each patient. The optical density versus the dose curve, followed the expected pattern, showing a good prediction from the General model, that the films employed in the exposures were of good quality and standard. Hence the optical density versus dose sensitometric curves depicts the outcome of the various films sensitivity after an exposure to the X-ray radiation through the patients.
References
-
NUCLEMED BV (2020). Darklight Duo Combination Sensitometer. NUCLMED BV, t Hoog Stuk 27, B – 8800 Roeselare, Belgium. Registered company BE 0435.002.834.
Google Scholar
1
-
N.A.M, van Bree, M.H Idzes, H.Huizenga, and B.J Mijnheer (1994). “Film dosimetry for radiotherapy treatment planning verification of a 6MV tangential breast irradiation, “Radiother, Oncol. 31, 251 – 255.
DOI |
Google Scholar
2
-
R.L. stern, B.A. Fraass, A. Gerhardsson, D.L McShan, and K.L. Lam (1992). “Generation and use of measurement – based 3 – D dose distribution for 3 – D dose calculation verification, “Med. Phys. 19, 165 – 173.
DOI |
Google Scholar
3
-
A Frazier, M.Du, J. Wong, F.Vicini, R. Taylor, C. Yu, R. Matter, A. Martinez, and D. Yan (1995). “Dosimetric evaluation of the confirmation of the multileaf collimator to irregularity shaped fields, “Int. J. Radiat. Oncol., Biol., Phys. 33, 1229 – 1238.
DOI |
Google Scholar
4
-
J.M. Galvin, A.R. Smith, R.D. Moeller, R.L. Goodman, W.D. Powlis, J. Rubenstein, L.J. Solin, B.Michael, M. Needham and C.J. Huntzinger (1992). “Evaluation of multileaf collimator de for a photon beam, “Int. J. Radiat. Oncol., Biol., Phys. 23, 789 – 801.
DOI |
Google Scholar
5
-
X. Wang, S. Spirou, T. Losasso, J. Stein, C.S. Chui and B. Mohan (1996). “Dosimetric verification of intensity – modulated fields,” Med. Phys. 23, 317 – 327.
DOI |
Google Scholar
6
-
L.J. Bos, C. Danciu, C. Cheng, M. J.P. Brugmans, A. Horst, A. Minken and B.J. Mijnheer (2002). Interinstitutional variations of sensitometric curves of radiographic dosimetric films. Med. Phys. 29(8), 1772 – 1780.
DOI |
Google Scholar
7
-
GAMMEX (2016). Film Densitometer Peripheral Technical manual Model 5336. GAMMEX Middleton, WI, USA.
Google Scholar
8
-
Scarlat F, Scarsoreamu A, Oane M, Mitru E, Badita E (2008). Determinationof Absorbed Dose Using a Dosimetric Film. IX Radiation Physics & Protection Conference, 15 – 19 November, Nasir City, Egypt, 313 – 321.
Google Scholar
9
-
Artur T. (2003). Determination of Absorbed X-ray Radiation Dose in X-ray Diagnostics and Imaging. J. Med. Sci. 1 (2): 31 – 36.
Google Scholar
10
-
NBIRR (Nigeria Basic Ionizing Radiation Regulation) (2003).
Google Scholar
11
-
ICRP (International Commission on Radiological Protection) (1996). Radiological Protection and Safety in Medicine, Publication 73 (Oxford and New York: Pergamon Press).
Google Scholar
12
Most read articles by the same author(s)
-
Oladotun A. Ojo,
Musibau A. Ibrahim,
Peter A. Oluwafisoye,
Dose Estimation of Absorbed X-ray Radiation to Safeguard Human DNA Alteration: A Quality Control and Quality Assurance Approach , European Journal of Medical and Health Sciences: Vol. 3 No. 3 (2021)